Laboratory Evaluation of the SIGMA-TRANSWAB® Transport System for the Recovery of Candida Species Using the Clinical and Laboratory Standards Institute (CLSI) Document M40-A2
Elcocks,
2020
Rising healthcare complications due to fungal infections increase the importance of efficient specimen collection and maintenance systems for correct identification and diagnosis. The CLSI M40-A2 protocol provides guidelines for laboratories assessing quality of medical transport devices, including swab transport systems (STS). This study assessed the efficiency of the SIGMA-TRANSWAB® foam and flock swab in recovering and maintaining viability of different Candida spp. including C. auris, in different test conditions. Both swab types recovered and maintained viability of all Candida spp. with greater CFU at room temperature after incubation (24 and 48 h) in comparison with swabs maintained at 4 °C.